USD
946.24 - 1246.24/pcs
Class B autoclave with capacity of up to 18 litres. It is ultra-quiet and really easy to use. Its automatic operating is configured with different sterilization programmes. With door locking to prevent it from opening accidentally.
Description
Autoclave 18 litres Class B with USB.
Reference
Printer
Programme liquids
922-STE-B18D
No (optional accessory, it is sold separately. If you wish to purchase it, look for the reference 922-PRINTER04).
Yes
Programmes:
Programmes
Temperature
Pressure
Sterilising time
Overall time
Type
Max load
Max load per tray
Solid
134 ºC
210 kPa
4 min
20-35 min
Solid unwrapped material
4.5 Kg
1.2 Kg
121 ºC
110 kPa
20 min
30-50 min
Wrapped
134 ºC
210 kPa
4 min
30-50 min
Solid unwrapped material
4.5 Kg
1.2 Kg
121 ºC
110 kPa
20 min
35-55 min
Solid or hollow material simple wrapped
3.5 Kg
1.1 Kg
Liquids
134ºC
210 kPa
10 min
30-55 min
Liquid
1 Kg
0.3 Kg
121ºC
110kPa
30 min
35-60 min
Textile
134 ºC
210 kPa
8 min
40-65 min
Porous unwrapped material
1 Kg
0.3 Kg
Porous material simple wrapped
0.80 Kg
0.25 Kg
Porous material double wrapped
0.6 Kg
0.2 Kg
121 ºC
110 kPa
30 min
45-75 min
Hollow material simple wrapped
3.5 Kg
1.25 Kg
Solid or hollow material double wrapped
1.5 Kg
0.6 Kg
Prion
134 ºC
210 kPa
18 min
45-70 min
Porous unwrapped material
1 Kg
0.3 Kg
Porous material simple wrapped
0.75 Kg
0.25 Kg
Porous material double wrapped
0.6 Kg
0.2 Kg
Hollow material simple wrapped
3.5 Kg
1 Kg
Solid or hollow material double wrapped
1.5 Kg
0.5 Kg
Test B&D
134 ºC
210 kPa
3.5 min
22-35 min
-
-
-
Test Helix
134 ºC
210 kPa
3.5 min
22-35 min
-
-
-
Vacuum test
-
-
-
15-20 min
-
-
-
The minimum time required autoclave to be ready for use after connecting is 5 minutes.
The max. temperature for the sterilisation cycle of 134 ° C is 137 ºC
The max. temperature for the sterilisation cycle of 121 ° C is 124 ° C
Features:
Fully automatic with pre-programmed sterilisation cycles.
Chamber and door in stainless steel with electro-polished finish.
High precision control system for perfect sterilisation results.
User friendly interface.
LCD screen displays information about the process continuously.
Open type water tank, this feature makes it easy water filling and cleaning the tank.
Optional printer (not included. Reference 922-PRINTER04).
Languages: Spanish, English, German, French, Italian, Dutch, Polish, Romanian, Czech, Latvian, Lithuanian, Hungarian and Chinese.
Safety devices:
Breaker.
Protect your computer from any possibility of failure in the power supply.
Action: Interrupt the power supply.
Thermal circuit breaker.
Protection against short circuits and possible overheating in the transformer primary network.
Action: Temporary interruption (until the temperature has dropped) in the winding.
Safety valve.
Protection against a possible increase in pressure within the chamber.
Action: Escape steam until the restoration of a normal pressure level.
Micro switch to control the state of the door.
Exercises control in adjusting the door to start a sterilisation cycle.
Action: Transmits signal to alert the incorrect position of the door.
Thermostat for heating elements in the steam generator.
Protection against possible overheating in the heating elements in the steam generator.
Action: interrupts the power supply to the heating elements in the steam generator.
Thermostat for heating elements in the chamber.
Protection against possible overheating in the heating elements in the chamber.
Action: interrupts the power supply to the heating elements in the chamber.
Auto leveling hydraulic system.
The hydraulic system pressure returns to its natural if the sterilization cycle is interrupted manually.
Action: the internal pressure is restored automatically.
Door Lock
Effective to prevent accidental opening of the door protection.
Action: Prevents door opening by a hydraulic locking system.
Technical features:
Nominal Voltage: 220-240V; 50/60 Hz; 10A
Chamber dimensions: 247 x 350 mm
Breaker: F16A / 400V
Nominal power: 2200VA
Sterilisation temperature: 121 ° C / 134 ° C
Capacity distilled water tank: 2.5 L approx.
Operating temperature: 5 ° C / 40 ° C
Operating relative humidity: Max. 80%, non-condensing
External dimensions: 480 x 455 x 600 mm
Net weight: 47 Kg
Max noise level: <70 dB
Operating atmospheric pressure: 76/106 kPa
Components:
Distilled water tank
LCD display
Control panel
Main switch
Drain valve for distilled and used water
USB port
Safety valve
Ventilation grid
Lock system
Power cable
Rating plate
Accessories included:
Steriliser
Three trays of instruments
Base for trays
Extraction handle
Tool for door adjustment
2 drain hoses
Door gasket
User manual
Optional accessories and spare parts:
Printer for autoclave (ref. 922-PRINTER04) and paper roll for autoclave printer (ref. 534-ROLLPAPER).
Autoclave process step by step:
1- Open the door, take the tools and accessories that are within the sterilisation chamber, remove the wrapper and clean them.2- Plug the power cord to a suitable power source.3- Plug the printer (optional accessory, it is sold separately. If you wish to purchase it, please look the reference 922-PRINTER04).4- Switch on: The main switch is located on the right bottom in the front panel.Once activated, the LCD screen is on, where basic operation icons of the steriliser appear, such as the position of the door, the water level, the sterilisation programme, date, time, etc.5- Fill tank with distilled water.Remove the top lid and fill, if you hear a signal, it means that you exceeded the permitted level of water in the tank, please stop immediately and drain the amount of water necessary for the signal off.6- Preparation of material to be sterilised:Place instruments that are of different materials in different trays, however take care to separate them before sterilisation.If the instruments are not stainless steel, put some sterilisation paper between the tray and the instruments.Do not allow direct contact between instruments of different materials.Verify that all instruments are in correct position.Place on tray instruments such as lenses, containers, tubes, etc. inverted, so that water does not stagnate within them.Do not overload the trays above their limit.Do not overlap the trays on top of each other, or put them in direct contact with the wall of the sterilisation chamber.Do not use metal clips or pins, this may impair the operation of the steriliser.7- Select the sterilisation programme.8- Start the sterilisation cycle.After selecting the sterilisation programme, insert instruments into the chamber through the trays, using the supplied plastic handle.Once the instruments are within the chamber, close the door.Caution: Be sure to close the door, otherwise the steriliser displays an alarm and can not start or stop the programme during the cycle.9. Start the sterilisation programme.Press the start button, then the steriliser start operating during approximately between 30-75 minutes.10-After the sterilisation cycle.After completion of the sterilisation cycle, the printer will print the full report automatically (optional accessory with reference 922-PRINTER04)After the pressure drops to zero, you can open the door and remove the instruments. The steriliser alerts by a beep once the sterilisation cycle is completed.Caution: Whenever you remove instruments, use the supplied plastic handle for supporting trays.
IMPORTANT: Do not reuse waste water.
Installation:
There must be at least 10 cm of free space around the sterilizer, 25 cm at the top, and enough room for the door to open.
The place where you use the steriliser must have adequate ventilation.
The steriliser must be placed on a stable surface and at an adequate height for handling.
Do not cover or block the side outlets or outlet grille air condenser fan.
Do not place the steriliser near water sources where it can be splashed.
Keep away from heat sources.
Example of a printed report:
* The printer is an optional accessory and is sold separately. If you wish to purchase it, please look for the reference 922-PRINTER04.
================================Programme: WRAPPEDTemperature: 134CPressure: 206.0 kPaDrying time: 08minSterilisation time: 4.0Min--------------------------------Time Temp. pressureStart 12:28:17 089.0CT1: 12:31:32 087.1C -075.0kPaT2: 12:33:43 110.2C 052.0kPaT3: 12:36:37 088.9C -080.0kPaT4: 12:39:20 114.7C 053.7kPaT5: 12:43:37 087.9C -080.0kPaT6: 12:50:40 134.8C 206.0kPaTS: 134.7C 209.5kPaMax. Temperature: 135.2CMin. Temperature: 134.3CMax. Pressure: 214.0kPaMin. Pressure: 204.9kPaT7: 12:54:39 134.4C 211.4kPaT8: 12:57:36 102.1C -060.0kPaT9: 12:59:54 098.2C -060.0kPa13:04:07 End 102.4C--------------------------------Cycle No .: 00017Result: successDate: 07/06/2014SN: E54723operator:================================
Regulations:
93/42/EEC Medical Device Directive.
97/23/EEC Pressure Equipment Directive.
EN 13060 Small steam water steriliser.
EN 61326, EN 61000 Electromagnetic compatibility.
EN 61010-1, UL 61010-1 Safety requirements.
EN 61010-2-040, IEC 61010-2-040 Specific requirements for steam sterilisers used for treatment of medical materials.
In the UK this product needs to be inspected and certified anually. BVM Medical offer this service We do however have a technical service based in Spain to cover the product during the warranty period