Autoclave Black Dragon 12 liters class B for tattoo studios.
Programmes:
Programmes
Temperature
Pressure
Sterilisation time
Total time
Type
Max. load
Max. load per tray
Solid
134 0C
210 kPa
4 min
14-30 min
Solid unwrapped
4 Kg
1.2 Kg
1210C
110 kPa
20 min
30-45 min
Wrapped
134 0C
210 kPa
4 min
30-45 min
Solid unwrapped
3.5 Kg
1.1 Kg
121 0C
110 kPa
20 min
35-55 min
Solid or hollow simple wrapped
3.5 Kg
1.1 Kg
Liquid
134 0C
210 kPa
10 min
30-55 min
Liquid
0.80 kg
0.25 kg
121 0C
110 kPa
30 min
35-60 min
Textile
134 0C
210 kPa
8 min
30-50 min
Porous unwrapped
1 Kg
0.3 Kg
Porous simple wrapped
0.75 Kg
0.2 Kg
121 0C
110 kPa
30 min
40-60 min
Porous double wrapped
0.5 Kg
0.15 Kg
Hollow simple wrapped
3.5 Kg
1 Kg
Solid or hollow double wrapped
1.5 Kg
0.5 Kg
Prion
134 0C
210 kPa
18 min
30-50 min
Porous unwrapped
0.1 Kg
0.3 Kg
Porous simple wrapped
0.75 Kg
0.2 Kg
Porous double wrapped
0.5 Kg
0.15 Kg
Hollow simple wrapped
3.5 Kg
1 Kg
Solid or hollow double wrapped
1.5 Kg
0.5 Kg
B&D test
134 0C
210 kPa
3.5 min
22-35 min
-
-
-
Helix test
134 0C
210 kPa
3.5 min
22-35 min
-
-
-
Vacuum test
-
-
-
18-20 min
-
-
-
Safety:
A cycle will not start unless the door is firmly closed.
An automated system with double hydraulic locking device prevents the door from opening while chamber is pressurized.
The safety valves prevent overpressure in the generator and in the chamber.
A safety thermostat prevents overheating in the chamber and in the generator.
An automatic safety lock prevents accidental opening of the door during the sterilisation. A power protector automatically cuts off power in case of short circuit or if power supply is not very stable.
The failure alarm indicates a cycle failure and provides an error code.
The door alarm indicates that the door is not properly closed.
The low water alarm indicates when the water tank is empty or too full.
Technical features:
Power supply: 220-240V; 50/60Hz; 10A
Chamber: 200 x 360 mm
Breaker: F16A / 400V
Nominal power: 2200VA
Sterilisation temperature: 121 0C / 1340C
Capacity of the distilled water tank: 2.5 L
Working temperature: 5 0C / 40 0C
Working relative humidity: Max. 80%, non condensing
Overall dimensions: 414 x 365 x 570 mm
Net weight: 51 Kg
Max. noise level: <70 dB
Atmospheric pressure: 76 / 106 kPa
Components:
Distilled water tank
LCD display
Control panel
Main power switch
Drain connector (distilled and used water tank)
USB port
Safety valve
Ventilation grid
Circuit breaker
Power cable
Rating plate
Accessories included:
Steriliser
USB
3 instrument trays
Tray rack
Tray holder
Door adjustment tool
2 draining hoses
Door gasket
User manual
Standards:
93/42/EEC Directive on Medical Devices.
97/23/EEC Directive on Pressure Equipment.
EN 13060 Small Steam Sterilisers.
EN 61326, EN 61000 Electromagnetic Compatibility.
EN 61010-1, UL 61010-1 Safety requirements.
EN 61010-2-040, IEC 61010-2-040 Specific requirements for steam sterilisers used for treatment of medical materials.
AUTOCLAVE - step by step:
Open the door, take the tools and accessories that are within the sterilisation chamber, remove the wrapper and clean them.
Connect the power cord to a suitable power source.
Connect the printer (optional. It is sold separately, ).
Switch on: The main switch is located on the right bottom area in the front panel.
Once activated, the LCD screen is on, where basic operation icons such as the position of the door, the water level, the sterilisation programme, date, time, etc. appear.
Fill tank with distilled water.
Remove the top lid and fill, if you hear a signal, it means that you exceeded the permitted level of water in the tank, please stop immediately and drain the amount of water necessary for the signal off.
Preparation of instruments:
Place instruments made of different materials in different trays, however take care to separate them before sterilisation.
If the instruments are not stainless steel, put some sterilisation paper between the tray and the instruments.
Do not allow direct contact between instruments of different materials.
Verify that all instruments are in correct position.
Place instruments such as lenses, containers, tubes, etc. inverted, so that water does not stagnate within them.
Do not overload the trays beyond their limit.
Do not overlap the trays on top of each other, or put them in direct contact with the wall of the sterilisation chamber.
Do not use metal clips or pins, this may impair the operation of the steriliser.
Wrap the instruments one by one. If you need to wrap several instruments together, make sure that they are made of the same material.
Seal the pouch with special adhesive tape for sterilisation or use a heat sealer.
Make sure that the pouch is properly positioned within the chamber on the trays (resting on the plastic part).
Choose the sterilisation programme.
Start the sterilisation cycle.
After selecting the sterilisation programme, insert instruments into the chamber on the trays, using the supplied plastic handle.
Once the instruments are placed within the chamber, close the door by turning the knob.
Caution: make sure to close the door, otherwise the steriliser displays an alarm and can not start or the programme stops during the cycle.
Start the sterilisation programme.
Press the start button, then the steriliser starts operating during approximately 30-75 minutes.
Sterilisation cycle finished.
After finishing the sterilisation cycle, the printer (not included*) will print the full report automatically and the data will be saved in the USB memory (.txt format).
After the pressure drops to zero, you can open the door and remove the instruments. The steriliser alerts by a beep once the sterilisation cycle is completed.
Caution: Use the supplied plastic handle for supporting trays when you take out the instruments.
IMPORTANT: Do not reuse used water.
Installation:
There must be at least 10 cm of free space around the steriliser, 25 cm at the top, and enough room to open the door.
The place where the steriliser is used must have proper ventilation.
The steriliser must be placed on a stable surface and at an appropriate height for handling.
Do not cover or block the lateral outlets or the outlet air grid of the condenser fan.
Do not place the steriliser near water sources.
Keep away from heat sources.
Example of printed report:
================================Programme: WRAPPED
Temperature: 134CPressure: 206.0 kPaDrying time: 08minSterilisation time: 4.0Min--------------------------------Time Temp. pressureStart 12:28:17 089.0CT1: 12:31:32 087.1C -075.0kPaT2: 12:33:43 110.2C 052.0kPaT3: 12:36:37 088.9C -080.0kPaT4: 12:39:20 114.7C 053.7kPaT5: 12:43:37 087.9C -080.0kPaT6: 12:50:40 134.8C 206.0kPaTS: 134.7C 209.5kPaMax. Temperature: 135.2CMin. Temperature: 134.3CMax. Pressure: 214.0kPaMin. Pressure: 204.9kPaT7: 12:54:39 134.4C 211.4kPaT8: 12:57:36 102.1C -060.0kPaT9: 12:59:54 098.2C -060.0kPaEnd 13:04:07 102.4C--------------------------------Cycle No .: 00017Result: successDate: 07/06/2014SN: E54723operator:================================